Laminar air flows into the workspace and provides absolutely clean work zone. Both exhaust and contaminated air pass through main and outer HEPA filters. HEPA prefilters installed under the working table. Protection of the repairer. 1,2,3 bio-contamination levels.
Material:
- Housing made out of galvanized steel
- Powder coated in standard RAL 9010
- Workspace made out of polished stainless steel (AISI 304)
- Vertical laminar air flow
- Air flow velocity: 0.25-0.45 m/s
Protection type:
- Safe operator working environment
- Protection of open pharmaceutical substances
- Protection of booth surroundings
Standards:
- Efficiency 99,95% MPPS according to EN 1822 standard
- ISO class 5 according to U.S.F.S. 209 E
Standard options:
- Microprocessor controlled parameters from control panel
- Work hours & UV light hours counter
- Electrical socket 230 V built inside the workspace
- Electrical lifting of front protective glass
- Bactericide UV sterilization light
Other options:
- Air, Gas, Water and/or Vacuum media connections
- Pressure gauge or pressure sensor
- Other dimensions on request
- Workspace made out of stainless steel (AISI 316)
- Satin finish on stainless steel